EMAIL

| CO2 Sensor CM1106-C | |
|---|---|
| Measurement range | 0-5% vol, 0-100% vol |
| Measurement method | NDIR |
| Accuracy | 0~1%: ≤±0.1%vol; 1~5%: ≤±(0.05%+5% of reading); 5%~full range: ≤±10% of reading |
| Warm-up | <10s |
| Response time | T90<25s at 20℃ |
| Working voltage | 3.3-6.0V |
| Working current | 55-60mA |
| Output voltage | 0.4-2V/TTL |
| Digital signal format | RS 232 data bit:8 ;stop bit:1; check bit: null |
| Baudrate | 9600 bps |
| Working temperature | -40℃ to 70℃ |
| Dimensions | ¢20*19mm (except pin) |
| Output pin | 5 pins |
| Life span | ≥ 10 years |
| Ex-proof Level | Ex ia IIC T4 Ga |
Note: 1. The measurement accuracy is defined as temperature 25±2℃ and atmospheric pressure 101.3kPa.
The above functions and specifications are subject to the official product technical specifications. For more technical information, please contact: 027-81628827 or info@gassensor.com.cn.